aquamax kf portable engl

aquamax kf portable engl

Measuring and controlling the amount of water in oils is of considerable economic importance to the industrial community, particularly to the crude oil and electric power industries. Water content determination by Karl Fischer titration is no longer restricted to being used by a chemist in a laboratory – now the technique is used by engineers, plant operators, tanker drivers, distribution/maintenance engineers and other non-laboratory personnel.Karl Fischer titrations can now be performed in many different locations such as the tailgate of trucks, mobile laboratories, offshore installations, tankers, engineering workshops and dockside cabins. The Aquamax KF Portable is a truly mobile version of the very successful Aquamax KF Coulometer which includes Low Drift Cell twin port glassware as standard. Also incorporates a carry handle and a protective sleeve for the titration vessel and electrodes. The unique low drift cell glassware design is by far the easiest to use and also the most robust. The electrode locking system allows the joints to seal completely, without the use of grease or PTFE sleeves, and provides improved baseline stability. Hassle free assembly and disassembly. This glassware enables users to transport the instrument with reagents already in the titration cell so that it can be used immediately on arrival.

FEATURES

  • Simple operation
  • 10 user programmable methods
  • 1 ppm/100 %
  • Results in ppm, mg/kg, % water, μg water
  • Multi language display & print out
  • Small footprint
  • Integral high speed printer
  • Integral battery
  • Fully portable
  • Low drift cell design
  • Results Manager software
  • Automatically compensated errors

Technical Specifications

Titration method: Coulometric Karl Fischer titration
Electrolysis control: Patented “ACE” control system Gb2370641
End point detection: AC polarisation
End point indication: Visual display/print out/acoustic beep
Titration vessel: Low Drift Cell design,
no grease or PTFE sleeves required
Measuring range: Possible 1 μg - 200 mg water
Typical 1 μg - 10 mg water
Water content: 1 ppm - 100 % water
Max. sensitivity: 0.1 μg
Max. titration speed: 2 mg per minute
Max. current: 350mA
Drift compensation: Automatically controlled
Precision: 10 - 100 µg ± 3 µg; 100 µg - 1 mg ± 3 µg, above 1 mg ± 0.3 %
Start delay time: 0 - 30 minutes, user selectable
End delay time O- 30 minutes, user selectable
Calculation modes: Weight/weight (W/w) (user programmable),
Weight/dilution ratio (W/K)
Volume/density (V/SG)
Volume/volume (V/v)
Display format: µg, mg/kg, ppm, %
Print format: µg, mg/kg, ppm,
Method storage: 10 user programmable methods
Sample ID number: User programmable
Stirrer speed: Microprocessor controlled
Languages: Multi languages – user selectable
Calendar/clock: Analysis time & date print out
Battery low indicator: Display & print out indication
Data outputs USB and RS232 ports
Removable data storage: Flash drive (memory stick)
Data entry 15 key touchpad
Data entry 15 key touchpad
Display: 40 character alphanumeric
backlight LCD
Printer: 42 character high speed thermal printer

TOP