aquamax pro oil engl

aquamax pro oil engl

The Aquamax KF PRO Oil is the perfect instrument to measure ppm water in oils and fuels without the worry of interference side reactions caused by additives or sulphur/ mercaptans. The unique “closed loop” principle means no additional carrier gas is necessary. Directly injecting the sample in to the oven means no blank value is required, making the Aquamax KF PRO Oil a truly accurate, trace level water in petroleum products titrator.Learn more about your oils by using the temperature ramping program. This unique ECH feature allows you to see all various types of bonded water, making the Aquamax KF Pro Oil the perfect tool in the R & D, Refining, Used Oil analysis laboratories and lubricant blending plants. Crude oils are also perfectly analysed as we have the ability to display both free and bonded water. All Aquamax KF PRO Oil parts are totally enclosed making this system completely safe and robust for use in the demanding petroleum industry. The ECH technique allows a very long reagent life, because its capacity can be used completely

FEATURES

  • Water extraction of the samples at temperatures 35 °C up to 250 °C, e. g.:
  • Constant temperature
  • Individually set up temperature programs
  • Freely selectable temperature ramp
  • Time-controlled temperature programs for step-by-step heating
  • Type of result: µg, µg/L, mg/L, mg/kg, ppm, %, by using the formula generator

Applications

  • Transformer oils
  • Hydraulic oils
  • Engine oils
  • Gear oils
  • Crude oils
  • Fuel oils
  • Gasoline
  • Diesel
  • A1 Jet fuel
  • Military jet fue
  • Bio fuel
  • Petroleum products
  • Insulating oils
  • Silicone oils
  • Lubricating oils
  • Biopetroleum

Advantages

  • Closed loop principle does not allow methanol to evaporate from the KF solvent
  • Reagent capacity is used completely
  • Additive and Sulphur side reactions minimized
  • By using of temperature programs it is possible to separate free and chemically bonded water
  • emperature ramping program allows you to distinguish between various types of bonded water

Technical Specifications

Measurement method: Coulometric Karl Fischer titration
Sample administration: Manually with syringe
Sample amount: 0.01 ... 20 m
Heating temperature: 35 ... 250 °C, isothermal or with temperature program
Blank value: 0 µg
Measuring range: 0.0001 ... 100 %
Power supply: 230 V/50 Hz; 115 V/60 Hz
Dimensions: 33 x 39 x 48 cm (W x D x H)
Weight: 17 kg