The petroleum flash point tester.
eraflash X is the future-proof replacement for ASTM D93 Pensky-Martens testers. Tailored for the petroleum industry all kinds of liquids can be tested over the wide temperature range of 18 °C to 370 °C using only 2 mL of sample. The full compliance with the latest flash point standard ASTM D7094 brings 100% safety, unrivaled speed as well as best in class precision to your laboratory.
Measurement procedure
Just four easy steps are required for a safe flash point testing at maximum precision:
- High throughput with the 10-position autosampler
- Inherently safe with no open flame
- Only a very low sample volume of 1-2 mL needed
- Takes up very little laboratory bench space
- Fully compliant with ASTM D6450, ASTM D7094, and IP620
The instrument now visualizes the measurement procedure on the screen showing the sample and oven temperature as well as pressure gauge displaying the pressure increase after ignition. After the measurement the flash point temperature is corrected for the atmospheric pressure and rounded to the next 0.5 °C. The result will be saved in the corresponding result file.
Technical Specifications
Test Methods | ASTM D7094, IP 620 |
Fuel Specifications | ASTM D396, D975, D2880, D3699, D7467 |
PBT – Peltier Boost Technology® | High speed heating and cooling |
CPT – Contamination Prevention Technology™ | Advanced electrode protection and self-cleaning ignition system to minimize cleaning and maintenance |
QuickCal™ | In-situ calibration with eralytics unique calibration cups without the need of dismantling the instrument |
Fuel Dilution Program | Automatic fuel dilution measurement for the analysis of used engine oils |
Temperature Range | 18 °C to 370 °C (64 °F to 689 °F) |
Temperature Stability | 0.1 °C (0.2 °F) |
Turnaround Time | up to 5 tests per hour |
Sample Volume | 2 mL |
Reproducibility | R = 3.2 °C for diesel fuel at 60 °C (ASTM D93: R = 4.3 °C) R = 5.6 °C for lube oil at 200 °C (ASTM D93: R = 14.2 °C) |
Result Database | Over 100 000 detailed test reports stored |
PC Software | ERASOFT RCS – Windows® software for convenient data transfer and result analysis |
Remote Control | Remote service capability via Ethernet |
Interfaces | Built-in PC with Ethernet, USB, and RS232 interfaces, Wi-Fi via USB dongle Direct LIMS connectivity via LAN and output to printer or PC |
Display | Industry proven 5.7“ multilingual color touchscreen |
Languages | English, German, French, Spanish, Japanese, Chinese; additional langages available upon request |
Power Requirements | Auto-switching 85 – 264 V, AC 47 – 63 Hz, max 150 W, Field application: optional 12 V DC adapter |
Operating Conditions | Temperature range 10°C to 40°C, Humidity up to 90% RH, non-condensing |
Dimensions & Weight | W x D x H = 19.6 x 20.4 x 31.7 cm (7.7 x 8.0 x 12.5 in); 7.2 kg (15.9 lbs) |