SLFA-60/6000 series for a new generation of petroleum products :
Compact & Robust Design
Both the SLFA-60 and 6000 series have a protective membrane covering the cell window to prevent the detector and X-ray tube from accidental sample leakage. This protective membrane unit is simple to assemble and replace.
Large Screen Display
The easy viewing monitor screen will display real-time data in ppmor %.
Turntable (SLFA-6800 only)
The turntable equipped with the SLFA-6800 has been newly devised. Newly designed stepper motor is more robust and faster between sampling.
Measurement Procedure
- 1. Pour the sample into the sample cell
- 2. Place the sample cell on the sample holder
- 3. Set to the measurement condition
- 4. Press the “MEAS” button to start the measurement
Calibration
New software allows the SLFA-60 to store 3 sets of calibration and the SLFA-6000 series 5 sets. It also allows the users to choose between linear or quadratic curves. Calibration curves are restored manually without having to recalibrate as long as the users save the coefficients value/calibration data.
Output Options :
Print out, USB memory stick and USB cables connected to PC are available as output options.
* Printer can be switched off if not used.
Output Options :
The users can control the contents/volume of the information on the printouts. There are 3 layout options. Printouts show sample ID, X-ray spectrum, date/time, measurement average value, standard deviation and calibration curve graph etc.,on 80 mm wide format for easy viewing.
USB Output
The turntable equipped with the SLFA-6800 has been newly devised. Newly designed stepper motor is more robust and faster between sampling.
Safety Features
In addition to our thorough X-ray leakage tests at the factory, the SLFA series are equipped with these safety functions:
Key Switch
Only when the key switch is in the “ON” position, X-rays are turned on.
Shield Shutter
Even if the lid is open, X-rays are continuously on. However, the shield shutter covers the X-ray and will prevent leakage that might be emitted.
Shield Shutter
Even if the lid is open, X-rays are continuously on. However, the shield shutter covers the X-ray and will prevent leakage that might be emitted.
Technical Specifications :