Sulfimax headspace module

Sulfimax headspace module

The determination of volatile hydrogen sulphide (H S) in solid 2 and pasty samples is now very easy. The selective H S analysis 2 system Sulfimax GX of the ECH is extended with a manual headspace module.
The sample is filled into a glass vial and heated in the headspace module. Depending on the sample type, the heating temperature can be varied from 30 to 180 °C. The volatile H S is transported 2 through the gas circulation to the sensor in the Sulfimax GX and measured there. Very low H S concentrations are detectable. 2 Sample preparation is not necessary.
Due to the short determination times, high sample throughputs are possible. The compact and robust device can be easily operated by anyone

FEATURES

  • Complete separation of H S from the sample 2 by effective gas extraction
  • Simple calibration
  • Software: simple, clear, intuitive
  • Dosing manually by syringe
  • Definition of own methods for device contro
  • · Typical measuring time 5 min (depending on sample properties)

Applications

  • solid samples, e. g. elementary sulphur, sludge, bitumen,
  • liquid samples such as wastewater with sludge particles,
  • pasty samples,
  • soil samples and waste

Advantages

  • Easy connection to the Sulfimax GX Lab and Go
  • No sample preparation necessary for solid samples
  • Very low H2S concentrations detectable
  • Manually operated and robust measuring system
  • Easy handling for everyone

Technical Specifications

Sample type: solid, pasy and high viscous
Sample volume: 0.01 ... 20 mL (g)
Heating temperature: 30 ... 180 °C
Temperature resolution: 0.1 °C
Heating method isothermally
Power supply: 230 V
Power input: 100 W
Dimensions:: 130 x 270 x 290 mm (W x H x D)
Weight: 5 kg