Titramax tan tbn engl

Titramax tan tbn engl

At the end of the measurement, results are shown in mg Br /100 g 2 (Bromine Index) or g Br /100 g (Bromine Number). 2 The value of Bromine Index or Bromine Number expresses the trace amount of unsaturated constituents of hydrocarbons and oils. It is determined by measuring of consumed bromine for cleavage of double or multiple bounds during the titration. The value are needed for determination of parameters for further processing of petroleum products. The Titramax VT BROMINE is conform to standards ASTM D 1159, ASTM D 2710 and ASTM D 5776. The measurement uses a potentiometric titration method in an anhydrous medium. The titration with titrant starts, once the sample is dosed into the reagent. The user has to enter the sample weight into the menu. The titration is performed automatically until the endpoint indication of measurement.

FEATURES

  • an automatic volumetric titrator with potentiometric pH indication and integrated temperature sensor
  • a titration vessel with stirrer unit
  • The determination of alkalinity value is based on
  • an acid-base-titration in an aqueous medium
  • a precise indication by a selective electrode, which is stable over long periods
  • Steps of the analysis are

Applications

  • petroleum distillates
  • kerosine
  • gasoline (including leaded, unleaded, oxygenated fuels)
  • distillates in the gas oil range
  • commercial propylene trimer and tetramer
  • butene dimer
  • mixed nonenes, octenes, heptenes
  • olefinfree hydrocarbons or mixtures

Advantages

  • · Complete measuring system for the determination of Bromine Index or Bromine Number
  • Fully-automatic volumetric titration
  • Precise adjustment of the titration parameters by control algorithms
  • Preset measurement method allows an immediate start
  • The result output can be adjusted to your needs by using a formula generator

Technical Specifications

Measurement method: Volumetric titration
Types of result: mg KOH/g oil or using the formula generator
Measuring range / Display resolution: 0.01 ... 250 mg KOH/g oil / 0.01 mg
Measurement range pH / mV: - 3.0 ... 18.00 / - 2000 ... 2000
Display resolution pH / mV: 0.001 / 0.1
Accuracy pH / mV (without sensor): ± 0.002 / 0.1 mV ± 1 digit
Accuracy µA (without sensor): ± 0.2 ± 1 digit
Measurement range temperature °C: 0 ... 100 µ
Accuracy µA (without sensor): 0.2 ± 1 digit µ
Amplifier input impedance: 13 1 > 1 · 10 ohms